GOSSEN Mueller & Weigert 39
Vibration and Mechanical Shock Resistance
Influence factors for vibration and shock have been set forth in DIN EN 60 051. Our mea-
suring instruments comply with these requirements and are available as follows (see
respective data sheet for individual availability):
Effects of Vibration and Shock
Unless otherwise specified, class 1 and higher measuring instruments and accessories
must withstand the following shock and vibration tests, to which they are subjected during
the course of type testing:
Vibration Test
Vibration testing must be performed with the following specifications:
Sweep frequency range:
10 Hz - 55 Hz - 10 Hz
Wave amplitude: 0.15 mm
(corresponds to 1.5 g
n
at 50 Hz)
Number of sweep cycles: 5
Sweep velocity:
1 octave per minute
The vibration plane is vertical and the measuring instrument is mounted to the vibration
table in its usual operating position.
Shock Test
The shock test must be performed with the following specifications:
Peak acceleration:
a) 147 m/s
2
(15 g
n
)
b) 490 m/s
2
(50 g
n
)
No further explanation is required for peak acceleration value a). The manufacturer
must indicate a peak acceleration value of 490 m/s
2
for value b).
Waveshape: semi-sinusoidal
Number of shocks: 3 shocks each in both directions for 3 mutually perpendicular axes
(a total of 18 shocks)
Shock duration: 11 ms
The measuring instrument must be secured such that one of the three axes corresponds
to the direction of motion of the rotational axis of the measuring mechanism.
After completion of this test, an additional measuring error of 100% of the error class
rating may not be exceeded.
Scale and Pointer Types
Scale and pointer types for quadrant and circular scales are defined in DIN 43 802, parts
2 through 4 (as of size 48 x 48 mm), as well as for horizontal and vertical scales (as of
size 48 x 24 mm).
Our square and rectangular panel meters with edgewise scale are in compliance with
these standards.
Unless otherwise specified, our measuring instruments and limit transducers
(MESSCONTACTER) are provided with the following degree of protection in accordance
with DIN VDE 0470:
IP 52 for housing front panel
IP 00 for terminals
Climate-Proof Measuring Instruments
In their “conditionally tropic-proof” versions, our measuring instruments are especially
well suited for:
Moisture endangered areas in the temperate zones
Indoor use in the dry tropics
Indoor use in the wet tropics, during which condensation or seepage water, which may
be caused by air-conditioning, must be avoided.
Mechanical Stress Shock Resistance Vibration Resistance
Standard model
15 g
n
11 ms
0.15 mm
5 ... 55 Hz
Heightened requirements,
LN56
30 g
n
11 ms
2.5 g
n
5 ... 55 Hz
Heightened requirements,
LN55
50 g
n
11 ms
5.0 g
n
5 ... 55 Hz
Factor Allowable Values
Standard Measuring
Instruments
Conditionally Tropic-Proof
Measuring Instruments
Operating Temperature
Range
-25 ... + 40 °C -25 ... + 55 °C
Relative Humidity max. 85 %
(max. temp.: + 27 °C)
not for more than 60 days
per year
remaining days: 75%
annual mean: 65%
max. 95 %
(max. temp.: + 25 °C)
not for more than 30 days
per year
remaining days: 85%
annual mean: 75%
Condensation none none
Universal Program, General Overview