Simple Weighing
Place your sample on the weighing pan (7), and close
the draft shield door. Read off the weight indicated
on the display (28) only after the weight unit (“g”, or a
different
unit selected – see part 2, “Balance
Operating Menu”)
appears as the stability symbol.
Taring
If you wish to use a container or if the weight display
does not indicate 0.000 mg (depending on the
display mode or weight unit selected), zero the display
before you weigh.
To do so, press t key (26).
During taring, you can have the fully automatic draft
shield function either on or off.
For more information on turning this fully automatic
draft shield function on or off by menu code, refer to
part 2, “Balance Operating Menu.
Important Note for Verified Balances Approved for
Use as Legal Measuring Instruments:
The small circle in the weight display (on the left)
shows that the balance is exactly tared to “0”
(± 0.25 of a scale interval).
Important Note Concerning Verified Balances
of Accuracy Class k
To avoid measuring errors, the respective air density
must be allowed for. The following formula is used to
calculate the mass of the sample:
1– ρ
L
/8000 kg m
–3
m = n
w
1– ρ
L
/ρ
m = mass of the sample
n
w
= weight readout
ρ
L
= air density during weighing
ρ = density of the sample
1–26